A permittivity measurement system for high frequency laboratories

dc.contributor.advisorPalmer, K. D.
dc.contributor.authorMarais, Johannes Izak Frederiken_ZA
dc.contributor.otherUniversity of Stellenbosch. Faculty of Engineering. Dept. of Electrical and Electronic Engineering.
dc.date.accessioned2008-02-06T09:26:30Zen_ZA
dc.date.accessioned2010-06-01T08:26:40Z
dc.date.available2008-02-06T09:26:30Zen_ZA
dc.date.available2010-06-01T08:26:40Z
dc.date.issued2006-12en_ZA
dc.descriptionThesis (MScEng (Electrical and Electronic Engineering))--University of Stellenbosch, 2006.
dc.description.abstractThe open-ended coaxial probe is revisited as a broadband measurement system for general high frequency permittivity measurements. Three coaxial probes were developed that are suited for the measurement of both liquids and solids. The components of a permittivity measurement system were investigated and improvements were made to the coaxial probe where needed. This includes the development of a full wave code with great calculation time improvements without sacrificing accuracy. This code allows measurements to be performed in a high frequency laboratory and the permittivity extracted without any mentionable delay. A capacitance model that better describes the impedance of an open-ended coaxial line is also suggested that can be used for real-time permittivity extraction over a limited frequency range. Calibration formed a vital part of the project and great time was spent developing a TRL and a SOLT calibration set for the coaxial probe geometry. The combination of the TRL and SOLT standards also allows measurement of the residual errors after calibration and is used in an uncertainty analysis of the extracted permittivity. Well known materials such as PTFE, PVC, methanol and water were measured to test the probes. The measured dielectric constants are all within 3% of values quoted in literature. The loss term of the samples are also in good agreement with the expected values.en_ZA
dc.format.extent7446248 bytesen_ZA
dc.format.mimetypeapplication/pdfen_ZA
dc.identifier.urihttp://hdl.handle.net/10019.1/1524
dc.language.isoenen_ZA
dc.publisherStellenbosch : University of Stellenbosch
dc.rights.holderUniversity of Stellenbosch
dc.subjectDielectricsen_ZA
dc.subjectDissertations -- Electronic engineeringen_ZA
dc.subjectTheses -- Electronic engineeringen_ZA
dc.subject.otherElectrical and Electronic Engineeringen_ZA
dc.titleA permittivity measurement system for high frequency laboratoriesen_ZA
dc.typeThesisen_ZA
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
marais_permittivity_2006.pdf
Size:
7.1 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.72 KB
Format:
Plain Text
Description: